Assembly of extracorporeal circuit with Efferon® device in isolated hemoperfusion mode on MultiFiltrate RRT machine or its analogues.
Plastic bag for priming solution, 2 liter
DIN-Luer bloodline,
2 pcs
0,9% saline solution bag, 1 liter
Heparin,
10.000 МЕ
Clamps,
4 pcs
Assemble the extracorporeal
circuit of the RRT machine
according to its user manual.
Stop the blood pump and put
clamps on both bloodlines
(“red” one is filled)
Add 10.000 ME heparin
to saline solution and mix
it to prepare the rinsing
solution
Connect DIN-Luer bloodline 1 with
priming solution bag. Fill the
bloodline with the priming solution
using gravity. Attach the clamp to the
bllodline.
Mount the Efferon device
in vertical position.**
Remove the cap from
bottom port of the Efferon®
device and connect the
pre-filled DIN-Luer bloodline
1 to it.
Remove the cap from upper
port of the Efferon® device and
connect the empty DIN-Luer
bloodline 2 to it.
Connect the plastic bag
for rinsing solution
to DIN-Luer bloodline 2
Remove the clamp from
DIN-Luer bloodline 1 and
rinse the Efferon® device
using gravity.
Put the clamp on DIN-Luer
bloodline 2 after the priming
is finished.
Disconnect the DIN-Luer
bloodline 1 from bottom port
of the Efferon® device
And connect the “red”
bloodline of extracorpore-
al circuit to it.
Disconnect the “priming”
bloodline DIN-Luer 2 from
upper port of the Efferon®
device
And connect the “blue”
bloodline of extracorporeal
circuit to it.
The circuit is ready. Remove
all the clamps and start the
blood pump
the extracorporeal treatment:
— 160-210 seconds of ACT
— 60-80 seconds of aPTT
device connection